Zeiss 710 MP microscope

 

If the startup gets stuck here for longer than a minute (a real minute, not just an misperceived eternity that feels like a minute), the most likely cause is the Real Time Controller.

 

To reset the RTC look under the computer table for the block box to the right of the computer.

Open the front panel.

Press ONCE only the bigger button labeled "RTC-RESET".

If the boot status slider does not progress after 2 real minutes, then turn off everything and try to restart.

 

AND CONTACT ME IMMEDIATELY TO TELL ME WHAT HAPPENED!

Michael.Cammer@med.nyu.edu or 914-309-3270 (by text if it's late at night!)

 

 

 

** Zeiss 710 MP instruction manual ** for basic powering up and powering down.